Every time a new ROI has been selected, one has to grab new references.

HeNe: Manual alignment procedure

  • Find reference in the laser alignment reference tab

Note the coordinates are for NEGATIVE, NEGATIVE which indicates the direction of the axes.

...

IR: Matching IR to HeNe beam path

Now after aligning the HeNe, we can proceed with using the IR laser and overlapping its beam path with the HeNe.

  • Check references on DMNear and Far for both pulsed and CW
  • Stop/Pause auto aligner

...

Image Removed

  • Close Iris near the DM to -5 mm (now motorised).
  • Increase exposure on B0 and B1 to 2 s (or see Laser cameras and alignment references)
  • Now someone has to bring on the laser in the tunnel (close laser curtains and turn key)
  • Open shutter on Laser → LI20→ Laser and then the shutter next to LSS EPS EPICS from High to Low

Image Removed

...

  • For main beam alignment, align now to references on CompNear/Far and PBNear/Far. Suitable iris setting is -10 mm.