  1. Collect data, about 4 scans, in accuracy 5, with Multiplier 3.
  2. Observe the pressure (sum of scanned masses) at the RGA head
  3. It should now be within a factor of 2 of the CCG
    1. If this is not the case, repeat the calibration process for the multipliers, increasing or decreasing the gain depending on if the pressure reads to low or too high, respectively
  4. Collect 2 hours of data (preferrably 12-24hrs), Accuracy 6, and report RGA data/status to Peter, and later to ServiceNow
  5. Insert Calibration result into Google spreadsheet under 'Cart Qual Status' and to its respective Cart Tab.

EXTRA HELPFUL INFORMATION: