Check charge injection algorithm for selected pixels

Code Block
Hart, Philip Adam <philiph@slac.stanford.edu>
Thu 9/24/2020 11:52 AM
To: O'Grady, Paul Christopher
Cc: Dubrovin, Mikhail
I'm working in the data frame, except I've flattened the array for some reason.  So the standard bad behavior can be seen in module 6, pixel 7777 in my ntuple:
>>> 96*4
384
>>> 7777/384
20
>>> 7777%384
97

so [6, 20, 97] and all around there.
I see somewhat ok behavior in [14, 20, 97].
[2, 20, 97] and [10, 20, 97] seem plain ok.
[6, 4, 117] looks bad.  [2, 4, 117] seems ok.
[10, 218, 53] is pathological in a check-mark way.
[2, 286, 207] is pathological in a check-mark way.

- Philip
=====
Hart, Philip Adam <philiph@slac.stanford.edu>
Fri 9/25/2020 10:38 AM
Here are some pixels which are problematic in (calibrated) AML-L in module 6 in both water and air scatter for the limited events I scanned in mfxp175 run 108 and mfxc00118 run 190, respectively:

module row col nBadWater nBadAir
6 175 172 21 150
6 176 107 116 174
6 156 191 25 174
6 268 177 161 150
6 175 161 25 155
6 95 191 21 108
6 93 69 183 508
6 175 180 30 202
6 175 185 24 149
6 175 186 27 196
6 168 191 23 159
- Philip

good pixel 2, 20, 97

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

bad pixel 6, 20, 97

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G20,97

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

pathological in a check-mark way 10, 218, 53

Image RemovedImage RemovedImage Removed

Image RemovedImage Removed

pathological in a check-mark way 2, 286, 207

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

BadWater 6, 175, 172

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G175,172

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

BadAir 6, 21, 150

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G21,150

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

BadWater 6, 168, 191

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G168,191

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

BadAir 6, 23, 159

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G23,159

Image RemovedImage RemovedImage Removed

Image RemovedImage RemovedImage Removed

Chi2

Commands to generate plots

bsub -o 2020-09-28-mfxc00118-r0172-offset-log06 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o work -p -C -i6

plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-H.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-M.dat

chi2 for panel 2 AHL-H, AHL-L, AML-M, AML-L

Image RemovedImage RemovedImage RemovedImage Removed

chi2 for panel 6  AHL-H, AHL-L, AML-M, AML-L

Image RemovedImage RemovedImage RemovedImage Removed

Summary

  • Minor detector damage is observed in a few constant types - in charge injection gains for FL, AHL-L, AML-L
  • Charge injection gains in stead of uniform constant gains can be used to eliminate this effect.

...