Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

It is possible that this shoulder is partly due to the K-beta emission of the material used in this particular experiment which happens to be copper (while the main peak is the K-alpha emission). After a quick loop-up, these values for copper are approximately 8040 keV for K-alpha and 8900 keV for K-beta which gives a 1.107 ratio of K-beta to K-alpha. If we look at the distribution of just the maximum energy pixel of each peak, as shown below, there is very visibly some type of peak on the higher energy side of the 1 pixel distribution. The main peak has a mean of 140.3 while the mean of the smaller peak is about 155. This gives a ratio of 1.104. The numbers used are approximations and the second peak is artificially shifted by the Gaussian from the first peak underlying it. So it is very likely that these two peaks are the K-alpha and K-beta lines of copper.

Image Modified

Closer Look at the Effect of Pixel Size

Since the square peak finder only find 4 pixel peaks, only the data for the V1 peak finder is shown below.  This data consists of 29,342 peaks found from the 13 different events.

PixelsPeaksMeanErrorRMSImage Modified
17164140.240.097.45
214926142.540.078.24
34354141.530.1711.44
42898146.580.189.70

...

As can be seen in the table above, there is a noticeable difference in the means of the energy distributions for each number of pixels which, ideally, shouldn't exist. Furthermore, the errors on the mean cannot account for this difference. One theory is that the larger peaks (the ones with more pixels) end up adding in more noise to the total energy of the photon thus shifting it to a slightly higher energy. If such is the case, one remedy would be to shift each distribution so that their peaks fall on the same bin. This was done by using the average of the four bins where the peaks existed as the new bin for the peaks. Below is the result compared to the normal distribution shown in the first graph on this page. As can be seen, the difference is very slight; the distribution is slightly sharper for the shifted data, but not by very much.

Image Modified

When fit with a Gaussian in the same manner as before, the results confirm this observation. Although there is a decrease in the width, it is very small. The noticeable change in the mean is most likely due to the choice of energy used to which the peaks were shifted.

...