Wed 6/1/2016 2:12 PM:
Attached, the latest metrology files for the 3 quads at MEC.
Quad0 and Quad2 metrologies should be applied for all experiments after 2016.02.16.
Quad 1 metrology should be applied for all experiments after 2016.04.21.
As always, please use only Mikhail’s curated version; Excel files are known to contain bugs.
Thanks,
Gabriel
------------
On Thu, Jun 2, 2016 at 9:42 AM Dubrovin, Mikhail:
Hi Gabriel,
Your file with optical measurements of MEC Quad1 from 
2015-04-21 has large Z coordinates even for the 1st point as shown below.
Point 6 with Z=-1095 is definitely problematic. Should we trust to these Z measurements?
Mikhail
------------
Thu 6/2/2016 9:47 AM
Hi Mikhail,
No, the very large numbers (in absolute values) are not correct; sometimes the OGP focuses on other surfaces than the sensor surface. I’ll look into the strange values today.
(If a large fraction of the tiles are not correct, we can easily re-metrologize the offending MEC quad; we only need to remove the front shield. In the CXI camera it’s much more complicated, as written yesterday).
Thanks,
Gabriel
Code Blocknote
Blaj, Gabriel Fri 6/17/2016 4:53 PM
To: Galtier, Eric Christophe;  Blaj, Gabriel; Dubrovin, Mikhail; Cc: Hart, Philip Adam;

Hi,
Just to make sure there is no labeling mix-up, could you please try the
latest metrology files for the other 2 quads? I’m asking this because 2
weeks ago I received a quad I thought was 1 but was labeled 2
Gabriel

...