ang brings us an image: - the same image, but spectrum is shown in the range (-100, 100) ADU. There is no peak at -35ADUThis event looks good.
Event #36: looks like left-side sensor is not corrected for common mode fluctuation...
In stead of default common_mode parameters (1,25,25,100) let us try (1,50,10,100) in the file
Code Block |
---|
/reg/d/psdm/CXI/cxi86415/calib/CsPad2x2::CalibV1/CxiDg3.0:Cspad2x2.0/common_mode/9-end.data |
The same image looks better now:
Recommendation: try to play with common mode parameters.