  • Optical Metrology file: 
  • 2014-09-24 -- fixed and moved to DS2 chamber
    • mixed quadrants: 
      • 3 old A boards (Q0, Q1, Q2), 4/40
      • 1 new board (Q3), 3/19
  • 2014-11-19 -- DS1 @ DSD
    • moved back to DS1 chamber in DSD location
    • Used for PCS, IH and cxig3616
  • 2015-01-10 --  DS1 @ DSC location
    • Used for cxig1515
      • src = 'DetInfo(CxiDs1.0:Cspad.0)'
      • alias = DscCsPad
      • Relative detector Z position:  
        • alias = 'DscCsPad_z' 
        • pv = 'CXI:DS1:MMS:06.RBV'
  • 2015-03-17 -- DS2 @ DSD location 
    • Camera 1 was moved to DS2 chamber at DSD location.
       
    • The previously broken D50 Concentrator was replaced with older concentrator.
      • Quadrants 2 and 0 were swapped.
    • Back detector (generally not used) for SC1 experiments from 2015-03-18 to 2015-05-04
      • src = 'DetInfo(CxiDs2.0:Cspad.0)'
      • alias = DsdCsPad
      • Relative detector Z position:  
        • alias = 'DsdCsPad_z' 
        • pv = 'CXI:DS2:MMS:06.RBV'
  • 2015-05-04 -- DS2 @ DSD 
    • Cables for Camera 1 in DS2 corrected and collimator pipe removed from front DS1
    • Back detector for SC1 experiments from 2015-05-04 to 2015-05-17
      • src = 'DetInfo(CxiDs2.0:Cspad.0)'
      • alias = DsdCsPad
      • Relative detector Z position:  
        • alias = 'DsdCsPad_z' 
        • pv = 'CXI:DS2:MMS:06.RBV'
  • 2015-12-07 -- DS1 @ DSC
    • Camera1 was moved from DS2 (at DSD) to DS1 (at DSC) because Camera2 in DS1 was damaged.
      • src = 'DetInfo(CxiDs1.0:Cspad.0)'
      • alias = DscCsPad
      • Relative detector Z position:  
        • alias = 'DscCsPad_z' 
        • pv = 'CXI:DS1:MMS:06.RBV'

Camera2_V16_20150120

  • 2015-02-02 -- DS2 chamber with refurbished Camera2 moved to DSA location
    • Non-standard configuration given D50 CsPad rack concentrator, which is normally used with DS2 chamber, was broken.
    • Used D51 CsPad rack for concentrator and motors Februatry 2015 experiments in the SC2 (a.k.a. SC01, 100nm focus chamber). 
      • cxig0715, cxie7015, cxie7515 and cxif5315
      • src = 'DetInfo(CxiDs2.0:Cspad.0)'
      • alias = DsaCsPad
      • Relative detector Z position:  
        • alias = 'DsaCsPad_z' 
        • pv = 'CXI:DS1:MMS:06.RBV'
      • Similarly, other relevant epicsStore aliases begin with DsaCsPad.
    • The mapping of the D50 concentrator is rotated by 90 relative to the D51 concentrator.
  • 2015-03-03 -- Moved DS2 chamber with Camera2 to DSD location
    • Used for SPI experiment cxi86415
      • src = 'DetInfo(CxiDs2.0:Cspad.0)'
      • alias = DsdCsPad
      • Relative detector Z position:  
        • alias = 'DsdCsPad_z' 
        • pv = 'CXI:DS2:MMS:06.RBV'
    • Note pipe in front DS1 chamber detector during this experiment was removed, but the holder for it was still in providing circular aperture for back DS2 detector
  • 2015-03-17 -- Added shield and pipe and moved Camera2 to DS1 chamber
    • Used for SC1 experiments starting at the end of Run10:

      • cxig0915, cxii6315, cxih0115, cxih0815, April PCS, cxih9515, cxii2415, ...
      • src = 'DetInfo(CxiDs1.0:Cspad.0)'
      • alias = DscCsPad
      • Relative detector Z position:  
        • alias = 'DscCsPad_z' 
        • pv = 'CXI:DS1:MMS:06.RBV'
    • Detector Attenuation from shields and coatings (pipe and additional shielding required for laser experiments):
      • CsPad Pipe Al Shield Transmission:
        • 8 um thick polyimide 

        • 1 um thick Al on upstream side

        • 1.75 um thick Al on downstream side

      • The detectors have 0.25 um thick Al coating.

      • --> 3 um total Al

  • 2015-05-04 -- Removed shield and pipe from Camera2 in DS1 chamber
    • Used for SC1 experiments starting with cxih8015:

      • src = 'DetInfo(CxiDs1.0:Cspad.0)'
      • alias = DscCsPad
      • Relative detector Z position:  
        • alias = 'DscCsPad_z' 
        • pv = 'CXI:DS1:MMS:06.RBV'

...