Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

There were no data links present. Looking at the
oscillator that is used to driver the links, the
frequency is at half of what is it suppose to be.

1/23/15 - TKN

Checked the grounding scheme of the detector.

L1-3 u-channel / SVT Box: 0.1 Ohms

FEB cooling plate / SVT Box: 6.7 Ohms

FEB cooling plate / table: 4.6 Ohms

SVT Box / table: 3.4 Ohms

Seems to imply that the table is between the FEB cooling plate and the SVT Box.  Perhaps implies a connection from FEB cooling plate to vacuum box (e.g. via data cable edge contact) and then vacuum box to table then table to SVT Box?

We should try to break this when we can.

 

Equipment

Interlock board

...