12:15
Continue to 100pC and 180V.
Here we see a drastic change and what looks like many shorts has developed across the sensors. We don't see any case where the channel get a larger noise but they all go down to a level of about 20ADC.
The step in noise also become worse.

Baseline taken at 180V, after irradiating with 100 pC at 180V:
Image Added

Baseline taken at 250V, after irradiating with 100 pC at 250V:
Image Added

Baseline taken at 180V, after irradiating with 100 pC at 250V:
Image Added

Baseline taken at 350V, after irradiating with 100 pC at 350V:
Image Added

Baseline taken at 180V, after irradiating with 100 pC at 350V:
Image Added

12:30
At 100pC and 350V.
Here we see a large increase in number of channels that are "shorted".
Interesting it seems like the behavior depends very sharply on the bias voltage setting. If we go to 210V the shorts are less pronounced but spike up above 215.
We don't see any real changes on the leakage current.

...