Operation
...
Running services
...
- If on, turn off NIM crate to disable interlock (otherwise chiller won't start)
- Turn on SVT power supply crate (turn key to "local on"), make sure red "interlock" light comes on (i.e. tripped)
- Set flow manifold valves
- Start chiller (power button on front panel)
- Turn on NIM crate, make sure Arduino's green "ON" light and flow switch's red "flow above threshold" light turn on, SVT power supply's red interlock light turns off
- Do stuff to start power and DAQ
...
Running tracker with CODA
...
- Login to tracker@ppa-pc88427 and open a terminal
- Setup environment: $cd /u1/software/daq && source setup_env.csh
- Kill processes related to the tracker and CODA (if any is running): $ cd /u1/software/scripts && StopLocalCodaAll
- On desktop #1:
- $ cd /u1/software/daq && RestartTrackerDaq
- $StartExpertGui
- Configure the SVT
...
Top half:
...
- using the Expert GUI:
- Hard Reset => Load Settings ("config/coda_defaults.xml")=> Load Settings ("config/coda_groupC.xml")=>Soft Reset
- You should see good sync (#apv's depends on what is connected)
- On desktop #2:
- open a terminal and setup environment: $cd /u1/software/daq && source setup_env.csh && cd /u1/software/scripts/
- $ ./StartLocalCodaAll
- You should see three terminals: Run Control Platform, coda master and trackerRoc and one coda master GUI
- Select "coda" in the coda master GUI => opens a new coda master gui where you need to enable buttons
- Click Event Builder in the coda master gui => should open a smaller gui
- Click Run Control in the coda master gui => should open the large run control gui
- Click: connect, Configure, Download, Prestart (check progress in trackerRoc (apv sync status) and in run control gui for each step).
- Run with Go.
- To look at the data being taken go to desktop #1:
- $cd /u1/software/daq && source setup_env.csh && StartOnlineGui
SVT Layout in Lab
Side in lab (from chiller side)
| Flange
| Cooling loop
|
---|
"away"
| | "away"
|
"same"
| | "same"
|
| | |
...
Systems
Power
CAEN SY1527
Serial console
...
Slot
| PS
| Type
|
---|
0
| A509
| DVDD for all hybrids
|
3
| A509H
| AVDD, V125 and bias
|
6
| A509H | AVDD, V125 and bias |
9
| A509H | AVDD, V125 and bias |
Notes/remember:
- A509 BIAS is not used
- A509H last two channels are not used for any hybrid: ch.07 and upwards on slot 9 in this setup.
Mapping between the SY527 slots and channels to hybrids:
- Details here (Test run version): half-module assignments - Power supply assignments updated.pdf .
Wiki Markup |
---|
*\[Update with correct slot/ch assignment valid for group C\]* |
...
- Remember that a connector is switched on the support plate (see table in the above link) when identifying layer with fpga and hybrid id on the DAQ side.
Cooling
Chiller
Polyscience 6506T
...
- Flow switch: FPR122 with low flow adapter (0.1-1.0 GPM) http://omega.com/Manuals/manualpdf/M1960.pdf
Typical flow:
Setting
| Flow (GPH)
| Pinlet (psi)
| Poutlet (psi)
|
---|
"Away" loop
| 22
| 30
| 10
|
"same" loop
| | | |
Both
| 25
| 26
| 16
|
Test of serial port (serial_hps_chiller.ino )
...