- Turn the gas ballast control knob of the scroll pump to level I
- If you are going to pump down within the next hour, stop here, otherwise continue the procedure
- Press STP next to the turbo pump logo
- Allow at least 15-20 minutes for the turbo pump to spin down (the off-status logo does not mean the blades have stopped spinning)
- Press CLS on SEW-VGC-15
- If you are going to pump down within the next 12 hours, stop here, otherwise continue the procedure
- Turn the gas ballast control knob of the scroll pump to level 0
- Open mechome, go to the tab "User Devices", open the panel "User Sentry PDU"
- Turn On "Roughing pump PTM 02 (turbo 2)"
Version tracking
Version | Release date | Description |
---|---|---|
v1.00 |
| Demonstration version |
v1.01 |
| Simplification of the procedure using only checklists |