Page History
...
New dark data collected in ueddaq02 to check improved detector performance
Run # | Gain modeComments | Gain range index | number of steps | |
---|---|---|---|---|
553 | FH,FM,FL,AHL,AML | 0,1,2,3,4 | 5 | -step|
554 | FL | grindex=2 | 1 | |
555 | FM | 1 | 1 | |
556 | FH | 0 | 1 | |
557 | AHL | 3 | 1 | |
558 | AML | 4 | 1 |
Dark run 553 processing results
...
- epixquad in runs 553-558 behaves much better than in runs 398, 401
- rms after dark processing of the run 353 553 are almost the same for all banks in all gain modes
- two threshold algorithm applied in run 401 should work in this case, but all of collected runs were dark, without signal.
- all images show damage of the detector in segment 1 near detector center.
...
Overview
Content Tools