New dark data collected in ueddaq02 to check improved detector performance

-step
Run #Gain modeCommentsGain range index

number of steps

553FH,FM,FL,AHL,AML0,1,2,3,45
554FLgrindex=21
555FM11
556FH01
557AHL31
558AML41

Dark run 553 processing results

RMS for different gain modes of runs 554-558

pedestals for different gain modes of runs 554-558

First 10 events det.raw.calib for different gain modes of runs 554-558

...

  • epixquad in runs 553-558 behaves much better than in runs 398, 401
  • rms after dark processing of the run 353 553 are almost the same for all banks in all gain modes
  • two threshold algorithm applied in run 401 should work in this case, but all of collected runs were dark, without signal.
  • all images show damage of the detector in segment 1 near detector center.

...