- Baseline has tails
- The difference between the mean baseline of sample 1 and the rest of the samples ranges from ~15-25 ADC Counts which is larger than expected (~8-10 ADC Counts)
- The noise is at levels similar to what was observed during the test run
- The noise across an APV peaks in the middle. The peak seems to be present in sample 1 only.
- The noise of the edge channels is ~180 ~190 ADC Counts which is unusually high.
...
Sample-to-Sample Shifts in the Noise
![](/download/attachments/170754717/rce0_hybrid0_run1_sts_var_noise.png?version=1&modificationDate=1401140022000&api=v2)
Channel Plots