This will open the device (xc7a200t_0). Right-click the device and choose Add configuration Configuration Memory device:

In the window that opens specify the memory device to be mt25ql256-spi-x1_x2_x4:

Image Added

Accept the prompt to start programming:

Image Added

In the window that opens up select which image you want to flash onto the detector (different for LCLS-I and LCLS-II):

Image Added

The image you want to specify for  LCLS-I can be downloaded here:

Epix100aGen2_EA020004.mcs

The one you want for LCLS-II can be downloaded here:

Epix100aGen2-0xEA020008-20211206091023-ruckman-cdd8531.mcs

After downloading the relevant image, give the path in the configuration file box:

Image Added

and press ok twice to start programming:

Image Added

After the flashing has been done, the following message should be seen:

Image Added

After this, close down the software, disconnect and reassemble detector and you are good to go.


The firmware images provided above is valid at the time of writing this page. If new version of the firmware becomes available the following page will instruct on how the get a hold of them, and also provides information such as how to check which firmware version is on the camera:


Firmware version tracking for ePix analog detectors