Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The low-Z XAS hutch configuration consists of two separate Helium gas spaces.  The front end / I0 section is supplied by SSRL house helium supply, while the low-Z sample box section is supplied by the Helium cylinder in the hallway behind the 4-3 blue electronics racks.  The two sections are separated by a thin polypropylene film inside a clamped O-ring assembly as indicated in the picture.

...

Remove one of the two paddles from the low-Z sample box and observe the flowmeter on the front/end I0 section.  If the flow drops to zero, then there is a leak between the two sections, and you will need to change the polypropylene film.

Image Added


To replace the polypropylene film and fix the leak between sections:

...

Test that the two gas sections are separated again.  To do this, check that the He flowmeter is at a high value (it reads "H +F" above 500) and decreasing.  The initial high value is because there is heavier gas (air) being purged out of the Front end / I0 section.  It should eventually settle around 200-240 when only Helium is in the front end / I0 section.  If the flowmeter value is already lower than this after clamping, this means there is a hole in the film.  This often happens because the O-ring can cut the film.  No worries!  Just unclamp the assembly, replace the film, clamp and test again!

...