Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

28 October 2022:

  • All detectors in this table are operational.
  • YAG cameras verified:
    • YAG0
    • YAG1
    • YAG3
    • YAG4 (new, by Zyla)

20 October 2022: Zyla update.

  • The camera was completely disconnected.
  • All connections remade except for the other end of the DB15 cable (power...)
    • Will look for this during housekeeping, searched multiple areas inn the hutch today.

19 October 2022: Resurvey status.

  • Xray_imaging and SPL_TC_alignment gigEs operating.
  • Verified operation of YAG1 and the SPL/Compressor gigEs (CompInNF/FF, CompOutNF/FF) not listed here.
  • All 4 Opals and the ePix100a at XTCS are running.
  • The Zyla is not appearing in the DAQ, will reset today.

The detector locations within the target chamber are marked on this diagram:

Image Removed

XRTSXTCSe- spectrometer

1st alignment camera

Do we still need this?
Detector NumberTypeFunctionIn DAQ?NotesCabled?Running?Comments
1ePix100asingle photon diagnosticsYesin-vacNW flangeYesYes

etch 12; vertical holder

Q0 fibers/PDU

ePix 0

Not usable at these intensities (>1019 ?)

2ePix100a

single photon diagnostics

Yesin-vacYesYes

etch 25; horizontal holder

Q1 fibers/PDUePix 1.

3ePix10kePix100aXRDXRTSYesin-vac4opal 2K

No


Yesin-airYesYesopal4_electron_spectrometer
5opal 4KThomson ParabolaYesin-airYesYes

opal3_thomson_parabola

Not used for L-10008.

6opal 2KNF/FFYesin-airYesYes

opal1_nf_ff

Replaced 1K with 2K.

7opal 4KImaging spotYesin-air

Yes

Yes

opal2_focus_img

8opal 4KTime ToolYesin-airYesYesopal7_timetool

ePix0

Q2 fibers/PDU

Recovering from water spill.

4


ePix100a

Kalpha

User spectrometer

Yesin-vacYesYes

Q3 fibers/PDU

59ZylaMXIYesin-airYesYes

zyla_xray_east

Restart procedure at Zyla and iStar confluence page.

Host is daq-mec-cam02.

106gigE POE

replacement for GIGE4

XTCAV
YesNoin-air, using old GIGE-MEC-CAM08 (Mako G-032B POE)vacYesYes

Now known as the beamline camera "XRay_Imaging"

11gigE POESPL_TC_alignmentNoin-air, using old GIGE-MEC-CAM07 (Mako G-032B POE)YesYesIt does run well. From time to time, it 'freezes' such that we have to start/stop again. Nothing serious. The camera works well for its purpose! 
12gigE POENoin-air, Mako G-032B POE, TBD for wire alignmentDo we still need this?13gigE POE

2nd alignment camera

Noin-air, Mako G-032B POE, TBD for wire alignmentDo we still need this?14contactless probe

alignment

Noin-vac, TBD, need only the laptop to be accessible remotely

ePix100a SPC vertical → Q0 fiber & PDU S0

ePix100a SPC horizontal → Q1 fiber & PDU S1

ePix100a XRTS → Q2 fiber & PDU S2

ePix100a Kalpha → Q3 fiber & PDU S3

Update on Fiber Mapping: (see 7.11 Opal Camera Configuration for fiber patch panel information)

  • Commenting out epix100a_2_XRTS and setting "-i 2" for epix100a_3_Kalpha worked (but keeping -P 0x40) worked, and I checked via image inspection that I saw two different detectors.
  • The epix100a persisted all day across nearly complete fiber changes - so the DAQ config is the issue (and the IOC was ok).
  • In detail: Dimitri and I got into the fiber routing behind the Quad connections in PR60 since we suspect cable problems coming into cam01. So now cam01 has both the direct fiber bundle and the Fiberstore QSFP-SR4-40G that was in one of the two pgp02 ports. We don't need the quads again until late Feb, so can sort this out post L10011.