Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Short Pulse Laser System

The short pulse laser at MEC is a double-CPA laser designed to operate at >25 TW peak power, with < 40 fs pulse duration and > 1 J per pulse. A schematic of the laser system is shown in Figure 1. A commercial CPA (Legend) system, seeded by an oscillator that is RF locked to LCLS, generates mJ-level compressed pulses, which are passed through an OPA pulse cleaning system before being stretched again for further amplification. The first bowtie amplifier operates at 120 Hz, matching the LCLS x-ray pulse rate, outputting >10 mJ per pulse. A second bowtie stage amplifies to >1.5 J at 5 Hz repetition rate. After compression, this yields > 1 J in 40 fs, for 25 TW peak power.

Image Added

Figure 1. Diagram of the MEC short pulse laser system.

Pulse Duration

The final output has a pulse duration of approximately 35-55 fs, with a beam diameter of 65 mm, and an energy of 1 J. The compressor grating spacing can be adjusted to produce between 35 fs and 8 ps pulses. For pulses of longer duration, stretched pulses with 150 ps duration have been provided successfully.

Image Added

Figure 2. (legacy information: please contact) Wizzler measurement of the pulse shape of the MEC short pulse laser indicating compression to nearly transform-limited pulse at 23 nm bandwidth.

Wavefront Correction

After amplification, a 90 mm clear aperture deformable mirror (DM; cf. Figure 3), using feedback from an online wave front sensor (WFS), corrects the distortions produced in the laser chain. Focal spot and wave front measurements in the chamber allow for pre-correction of wave front distortion from the compressor and focusing optics. This system is necessary for producing the highest intensity at focus.

Image Added

Figure 3. A photo of the deformable mirror of the MEC short pulse laser is shown.


Before CorrectionBefore CorrectionAfter Correction
RMS0.6 µm 22 nm
P-V2.5 µm 100 nm



Table 3. Demonstrated wavefront correction of the DM/WFS closed-loop system

Focusing

Focusing of the short pulse laser is achieved with an off-axis parabolic mirror. In table 4 below, OAP3 represents the optic used in the standard beam delivery for full intensity. See the description of the beam delivery on the MEC Components page of the LCLS website (until the Confluence page is updated).  OAP1 and OAP2 use narrow angles and are thus difficult to use with most target frames. OAP4 has a smaller clear aperture and will clip the full energy beam.

NameCA (mm)Effective focal length(mm)Off-axis angle (deg)coatingnotes
OAP199311.6617.55DielectricIncompatible with standard target mount
OAP299311.6617.55DielectricIncompatible with standard target mount
OAP399330.435.31SilverPrimary OAP for full intensity
OAP450500.5834.41SilverFor long focal length and low power. Only compatible with reduced beam size and power.

Table 4. The OAPs owned by MEC. OAP 3 is standard for full intensity experiments. OAP 1 and 2 have too tight of an angle to use with the standard target frame, but may be used in certain customized experiments. OAP 4 is intended for frequency doubled or uncompressed experiment in which the beam size is smaller than standard. Typical use is to produce an f/10 focus.


Figure 4. A false color log scale focal spot image of the focus from OAP with full wavefront correction. The measurement was taken at reduced power but full power inferred intensity is shown. Lineouts are shown in linear scale. The right image depicts a false color image in linear scale.

Radiation Interlock System

To prevent the exposure of personnel to ionizing radiation generated through high-intensity laser-matter interactions, the MEC SPL employs a Radiation Interlock System as a safety control. More information can be found here: 0.3. Radiation Interlock System.

Pulse Contrast

Nearly all user experiments using the multi-TW beam require excellent contrast to be successful. To achieve highest baseline pulse contrast, the short pulse laser system employs a double-CPA front end. After initial chirped pulse amplification of the oscillator pulses at 120 Hz, the beam is recompressed and passed through a nonlinear filter (with an OPA and SFG-based pulse cleaner [1] replacing XPW starting in Run 17) and then stretched again for final amplification. This filter removes prepulses and pedestal energy that builds up primarily in the laser front end.  Measurements utilizing third order autocorrelation techniques revealed a contrast better than 107 in the > 3 picosecond range using a double CPA pulse cleaning system. To further increase this contrast, MEC can provide a 2nd harmonic stage that, in combination with dichroics, should theoretically achieve better than 1014 contrast beyond 3 ps, while sacrificing pulse energy (output ~300 mJ at 400 nm).

Image Added

Timing Jitter and Synchronization with the LCLS X-ray Beam

Taking full advantage of the temporal resolution of femtosecond X-ray pulses and femtosecond optical lasers in pump-probe experiments requires timing measurement of a few 10 fs or better. LCLS has a pulse-to-pulse timing jitter relative to the accelerator radio-frequency (RF) distribution of approximately 60 fs RMS, integrated over a bandwidth of 0.1-100 kHz. The MEC short pulse laser oscillator is locked to the accelerator RF, with a distribution with similar or better timing jitter. Drifts in the laser beam path and RF distribution need to be controlled to approximately the 1 ps level.

The mode-locked seed laser oscillator operates at 68 MHz, the seventh sub-harmonic of the 476 MHz RF reference frequency of LCLS, which defines the fine (sub-picosecond) timing of laser pulses in the absence of configuration changes to the laser system.

The noise performance of the LCLS laser locking system was measured to have an RMS laser-to-RF-reference timing jitter of 25 fs between 100 Hz and 100 kHz. Below 100 Hz, phase noise is dominated by the noise of the linac RF reference.

Time Tool

To overcome the limitations posed by timing uncertainties, experiments involving femtosecond physical phenomena, such as transitions of core-shell electrons, employ X-ray/optical cross-correlation techniques to measure the relative timing between the laser and X-rays on a shot-by-shot basis. This technique, referred to as the time tool and depicted in Figure 5, aims to determine the timing information precisely. By utilizing ultrafast ionization induced by the X-ray beam in YAG windows, synchronization with the MEC laser can be accurately established. The ionization process renders the window opaque, enabling t0 (the edge of before/after) timing observations. The time tool employs an optical probe split from MPA1 with 1mJ and 120Hz rep-rate, routing through the designated compressor, delay lines, and telescope cross-correlating with the x-ray beam on a YAG crystal. To reference the time tool to the target plane, another YAG window is positioned at the laser focus location, and a small aperture is temporarily inserted to spread the laser focus at the target plane. The time tool exhibits an RMS jitter of approximately 100 femtoseconds between the MEC laser and the X-ray beam.

Image AddedImage Added

Figure 5. Image of the MEC timing tool.

Image AddedImage Added



Figure 6. Synchronization of laser to the X-ray the time tool. Left: The geometry of time tool employs x-ray and probe laser on a YAG crystal. Right/Top: laser shadography of a target window on two consecutive laser shots, arriving just before (left) and just after (right) an X-ray pulse (square-shaped beam); Right/bottom: The time tool shows arrival time of the ionizing X-ray pulse inline, demonstrating that the synchronization jitter between the two consecutive pulses shown here is less than 100 fs.

Diagnostics for the short pulse laser system

The short pulse laser diagnostic suite includes a variety of devices that are accessible on-line and on demand. Here is a list of the common diagnostics already installed in the toolbox.

  • On-shot diagnostics:
    • Time tool (synchronization)
    • Pulse energy
    • Near-field beam profile
    • Wavefront
    • Equivalent plane focal spot
    • Spectrum
  • On-demand diagnostics (off-line)​
    • ​SPIDER
    • 3rd order autocorrelation
    • Far-field beam profile
    • Motorized power meters

[1] “Pulse contrast enhancement via non-collinear sum-frequency generation with the signal and idler of an optical parametric amplifier,” vol. 114, no. 22, p. 221106, Jun. 2019.