Runs of sxrg3715 have image every 10 sec it means that 1199 events are empty. In order to get calibration files the calibman default parameters needs to be changed.

Start calibman, click on tab "Configuration" then on tab "Parameters" and set field "end:" to 100000 and field "scan" to "2000". After that regular dark run calibration should work. Do not forget to coise the detector ANDOR.

The same trick with event numbers can be acheived in the calibrun command skeeping empty 1195 events:

calibrun -e sxrg3715 -d ANDOR -c ./calib -P -D -r 46 -n 100000 -s 1195

or

calibrun -e sxrg3715 -d ANDOR -P -D -r 46 -n 100000 -s 1195

or

calibrun -e sxrg3715 -d ANDOR -P -D -r 46 -n 100000 -m 2000

 

Type common_mode can mode constantscan be copied from file with parameters for mode 2 (see : Common mode correction algorithms ), for example

...