Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Amplitude spectra with corrected gain

Run 901, trans.=0.002

  • Spectrum is shown for events 1,11,21,31, and 41, respectively.
  • The gain correction factor is set to "0" for bad pixels. This explains the one-bin peak at "0" amplitude.

Run 902, trans.=0.005

  • Spectrum is shown for events 1,11,21,31, and 41, respectively.
  • The gain correction factor is set to "0" for bad pixels. This explains the one-bin peak at "0" amplitude.

Conclusion

  • Gain factors have the RMS spread ~5-10% depending on high/low-gain mode and intensity of illumination
  • There is no significant improvements of the amplitude spectrum after the gain correction is applied