Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Run numbers are shown in histogram titles:

Bad pixels are removed. Top-left plot has a peak at "0", which arose in the difference of two ~820x820 plots from regions between ASICs.
The difference between dark image amplitudes at High and Low gain mode is ~155 ADC units.
Comparison of runs 899 and 904 (transm.=0.1) gives the gain ratio High/Low = 495/66 = 7.5
Comparison of runs 902 and 903 (transm.=0.05) gives the gain ratio High/Low = 26/4.5 = 5.8. So, the gain is not a lilear function of intensity...

Gain Correction Factors

Run 901 transm=0.002

Image Added

Run 902 transm=0.005

Image Added

Run 903(low gain) transm=0.005

Image Added

Run 904(low gain) transm=0.1

Image Added

Run 905(low gain) transm=0.5

Image Added

Run 906(low gain) transm=1

Image Added