- Measure the laser energy and update S10 energy document: https://docs.google.com/spreadsheets/d/1OkifFC2Dw0UpbIavq1T5kBR-vykIuglx/edit#gid=698192852
- Shutdown the Evolutions
- Turn off the pump laser using the computer control panel (red STOP button)
- Turn the key on the control units to "off"
- On the control panel select "ramp down the LBO"
- Wait until LBO temp reaches set point (70 75 degrees)
- Switch off control box
- Shutdown the Revolution (regen)
- Turn off the pump laser using the computer control panel (red STOP button)
- Turn the key on the control unit to "off"
- On the control panel select "ramp down the LBO"
- Wait until LBO temp reaches set point (70 75 degrees)
- Switch off control box
- Shutdown the Vitara
- Disable the RF lock
- Disable RF Lock
- Make sure to power down the piezo amp
- Turn off the pump laser using the computer control panel. Turn off "Diode On". Turn off "Remote".
- Turn the key to the control unit to "off"
- Disable the RF lock
...