Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. If it was open, close angle valve VV3202 VV3201 isolating the plasma beamline from the plasma turbo
  2. Close valve FV-PB on the foreline of the Picnic Basket. Turn off Picnic Basket turbo. Wait for 5 min for turbo to stop (the speed readback immediately goes to zero, but it takes some time for the turbo to stop)
  3. From the E300 panel - Set Fill Gas #1 to ON
  4. From the MFC controller - Adjust the "Flow setpoint" on the MFC to reach the desired fill pressure.
    1. There is a slow response so make changes slowly.
    2. In previous tests these set points were used with the MFC:
      1. Helium at 5 Torr = 420
      2. Argon at 5 Torr = 130
      3. Hydrogen at 5 Torr = 580
  5. It will likely take 5-10 minutes for the pressure to approach the pressure setpoint. Make small changes and wait for 5-10 minutes for the pressure to stabilize before proceeding with more adjustments.
  6. Continue to monitor the IP pressure and make minor adjustments to the flow rate if necessary.
  7. To adjust the pressure, simply increase or lower the setpoint on the MFC.

...