- logon to the rayonix machine from mfx-daq
ssh hsuser@con-ics-mfx-rayonix (password is hsuser) - kill the capxure process (assuming it's running. ps -auwx |grep capxure). If there's no capxure process running, don't worry about it.
- killall procServ (this kills any startDaqInterface_newcraydl processes that maybe running)
- cd ~/slac/, and do:
- ./startDaqInterface_newcraydl
- Set DAQ evr to run at 30Hz.
- In EVR, The Rayonix pulse delay should be 0.006, and the pulse width should be 0.0005. Without these settings, you might see distorted images.
- Run
https://www.rayonix.com/product/mx340-hs/
...