Place to find out where to find all controls related things related to LY59- Kirian Experiment.

Table of Contents

How to get F-12/ beam stopper to work on the machine (recommendation to do this before the start of the shift)

Workaround for when encountering F-12 and F-11(enables beam stop in XRT) Keyboard shortcuts not working 

    • Restart the gnome shell by pressing, ALT + F2 , typing "r" in the Enter command window. 
    • Reapply the gnome settings by running "gnome-settings.sh"  
    • This seems to be an issue with the red hat and has been observed in other hutches as well. SLAC IT has an open ticket with red hat for the same. Moving this to resolved for now CXI Specific issue. 

To open CXI Home Screen:

Type "cxihome" in terminal. 

Sample Delivery

On cxihome go to User Devices Tab

Image Added

The box highlighted in red contains most of sample delivery screens needed for Kirian. 

Mesh Devices -

Sc2, will bring up the Input and Output voltage for configured mesh device, note that the scale factor is hardcoded as 1 on the screen. But the values can be changed via hutc-python regardless. 

Acromag- 

Acromag Channel # 15 (when starting the numbering from 0) is setup as the Input/Output for the power supply. Acromag screen can be opened from there and monitored as well. 

Pegasus sample Injector Motors - 

On CXI Home Click on the button highlighted in red. 

Image Added

wait for the screen to open on the bottom-right-hand-side of the screen : 

Image Added

Note: Fine Y doesn't seem to be able to lift up or put down the load in vacuum, works fine in air. Ray is aware of this. Run current is at max, it is possible that motor isnt getting enough torque to handle the load. 

Sample Chamber Motors: 

VT Stepper stages/motors - 

On CXI Home Click on the button highlighted in red. 

Image Added

sample Fixed target X/Y/Z and Rotation Piezo motor are on this screen -  

Image Added

On CXI Home Click on the button highlighted in red : 

Image Added

Objective Lens X/Y/Z and Questar X and Y can be found on this screen - 

Image Added

Piezo stages/motors  (Slits and Apertures)- 

On CXI Home Click on the button highlighted in red : 

Image Added

Important note: CXI:USR:MMC:09 (Slit Y+) is not working. 

Image AddedImage Added