Use this guide to trouble shoot faults with the DPS: DPS Pump Troubleshooting
Test results: DPS Stability and Failure Modes
Summary of 3 day test of stability with the manual flow control, and reaction of the IP pressure levels to various DPS failure modes.
dstorey - 2022-12-14 - DPS Stability and Failure Mode Tests.pdf
Info |
---|
- The DPS watcher should behave normally under a static fill of the IP.
- If filling with hydrogen gas then the purge gas should be used on stages 1 and 2. If you don't, the foreline pressure will build up slowly and eventually impact beamline pressures.
- Open the valve on the 6 pack and ensure regulator reads about 10 psig
- Open the purge gas valves on US1 and DS1 - Channels 1 and 2 on Beckhoff controller "N".
- Be sure to stop the purge gas and close the valves when finished.
|
...