  1. Close angle valve VV3202 isolating the plasma beamline from the plasma turbo
  2. Close valve FV-PB on the foreline of the Picnic Basket. Turn off Picnic Basket turbo
  3. Close Beryllium window valve VV3208
  4. Adjust the flow regulator to reach the desired fill pressure. There is a slow so make changes slowly.
    1. In previous tests a setting of ~0.8 SCFH results ~5 Torr for He
  5. It will likely take 5-10 minutes for the pressure to approach the pressure setpoint. Make small changes and wait for 5-10 minutes for the pressure to stabilize before proceeding with more adjustments.
  6. Continue to monitor the IP pressure and make minor adjustments to the flow rate if necessary.

...


Drain the

...

IP:

  1. If using the Li oven, follow appropriate procedures to shut down and secure the oven
  2. When ready to drain the He gas - close the fill valve. The IP pressure will drop at an initial rate of ~1 Torr/min.
  3. Open valve FV-PB on the foreline of the Picnic Basket to aid in the pump down
  4. Start the PB turbo when the IP pressure is below 1 Torr.
  5. Close the valve on the gas bottle.
  6. Open the DS Beryllium window valve VV3208 after the IP pressure is <<1 Torr.

...