  1. Measure the laser energy and update S10 energy document: https://docs.google.com/spreadsheets/d/1OkifFC2Dw0UpbIavq1T5kBR-vykIuglx/edit#gid=698192852
  2. Shutdown the Evolutions
    1. Turn off the pump laser using the computer control panel (red STOP button)
    2. Turn the key on the control units to "off"
    3. On the control panel select "ramp down the LBO"
    4. Wait until LBO temp reaches set point (70 degrees)
    5. Switch off control box
  3. Shutdown the Revolution (regen)
    1. Turn off the pump laser using the computer control panel (red STOP button)
    2. Turn the key on the control unit to "off"
    3. On the control panel select "ramp down the LBO"
    4. Wait until LBO temp reaches set point (70 degrees)
    5. Switch off control box
  4. Shutdown the Vitara
    1. Disable the RF lock
      1. Disable RF Lock
      2. Make sure to power down the piezo amp
    2. Turn off the pump laser using the computer control panel. Turn off "remote controlDiode On". Turn off "Remote".
    3. Turn the key to the control unit to "off"

...