SymptomPotential IssueWhat to do
Roughing pump tripped offRadiation related faultSee - Resetting a roughing pump fault
Roughing pump won't startError message is present

Check "More..." status to see what error is listed.
See - Resetting a roughing pump fault

Roughing pump won't reset an errorThere is a real problem, or pump has been damagedSee - Resetting a roughing pump fault

Turbopump tripped off

Most likely related to loss of backing pressureFix the problem with the roughing pump/foreline configuration then restart the turbo as described in Restarting DPS pumps.

Radiation related faultSee - Resetting a turbopump fault
Roughing pump or turbo not responsivePump is in local mode, not remotePut pump into remote mode. See Setting pumps to remote mode
US4 stage cannot be restartedProbably radiation damageSee - Pump redundancy
Foreline valve won't openForeline interlocks need to be clearedReset foreline inerlocks interlocks from "Foreline Vlv Intlks..." panel






...

The roughing pumps have handheld controllers in the S20 FKG20-23 24 rack, second from the top shelf, sitting on the "CORNING" box:in the back of the rack, on the lower shelf:

Image AddedImage Removed

And they look like this:

...