Check for your problem in the table below, then see full instructions below that for how to fix the issue.

Troubleshooting guide:

SymptomPotential IssueWhat to do
Roughing pump tripped offRadiation related faultSee -
Restarting
Resetting a roughing pump fault
Roughing pump won't startError message is present

Check "More..." status to see what error is listed.
See - Resetting a roughing pump fault

Roughing pump won't reset an errorThere is a real problem, or pump has been damagedSee - Resetting a roughing pump fault

Turbopump tripped off

Radiation related faultSee - Resetting a turbopump fault
US4 stage cannot be restartedProbably radiation damageSee - Pump redundancy
Foreline valve won't openForeline interlocks need to be clearedReset foreline inerlocks from "Foreline Vlv Intlks..." panel








Resetting a roughing pump fault

This may be indicated by one or more of the following:

  1. The roughing pump is stopped (red) or disconnected (purple), or
  2. The foreline vacuum gauge is reading a value >1 Tor, or
  3. A warning or alarm is present in the "More..." panel of the roughing pump panel, or
  4. Something else that I have not though of.








In the case of pressure rise on gauges 3259 and 3291 due to a failure of one or more pumps in the DS-DPS system, this procedure should be used to isolate the beamline and stop the pressure from rising further.  See the attached document for the full procedure.

...