June 10, 2022 dan morton, larry ruckman, matt weaver, nick waters, zach lentz, cpo, chris ford, mona

  • dan has been interfacing with axilon (cologne, germany)
  • stick with renishaw
  • axilon designs brackets and cable routing and provides renishaw (
  • digitized signal comes out
  • lead times long
  • encoder is in the FEE (just east of entrance on massive granite)
  • need Larry/Matt/TID to do the custom timestamping: pcie with FMC and SFP for timing system
  • what are the cable-length constraints?  maybe do a wave8-style box or a dev board (only one of these at the moment?)?  prefer dev board at the moment. comes out as PGP to KCU1500 in the DRP.
  • do we need a second controls interface to this box?
    • requires an "IOC" (more work for controls group)
    • may make it easier for the scientists to use
  • need to be able to reset the counter to zero (to get absolute-encoder style behavior)
  • dan says MHz is overkill, but psana needs MHz.
  • may need errors sent back to DAQ
  • need to decide on a table-top model for TID prototype. looking at page 14 of the manual (and page 7, which shows max speed vs. clock and resolution)
    • Interpolation factor: RIX would like 1nm (20KD).  may decrease range?
    • Alarm format and conditions: A (line driven E output; all alarms)
    • Clocked output option: 50MHz
    • Options: A (This is what axilon suggested, Larry thinks he can adapt)
  • cable going into Larry's box will be as on page 5 of the manual
  • cable lengths on page 9 look good.
  • diagram on page 9 of the manual shows the analog signal getting converted to digital by TONIC interface.  Larry needs to supply 5V.
  • schedule (need to ask Georgi). mechanical install in January 2023.
  • account numbers: ask Georgi

...