- Collect data, about 4 scans, in accuracy 5, with Multiplier 3.
- Observe the pressure (sum of scanned masses) at the RGA head
- It should now be within a factor of 2 of the CCG
- If this is not the case, repeat the calibration process for the multipliers, increasing or decreasing the gain depending on if the pressure reads to low or too high, respectively
- Collect 2 hours of data (preferrably 12-24hrs), Accuracy 6, and report RGA data/status to Peter, and later to ServiceNow
- Insert Calibration result into Google spreadsheet under 'Cart Qual Status' and to its respective Cart Tab.
...