The beam has a ~10-13 mm diameter (will depend on final collimation) and is uncompressed , so at around 200 ps pulse duration.

The potential fluence on the mirrors can be in excess of 1 J/cm^2.

Pre-2022 RP attenuator

Currently installed RP attenuator.

...

  • Appears to introduce astigmatism
  • Ghosts when changing polarisation
  • Loads of different optics (splitting ratio and polarisation)


Image Modified

Fig 1: Currently installed RP-attenuator. Red arrows indicate the beam path.

...

Attenuation should be down to 0.35^4 = 0.015 (1.5%) or 800 mJ * 0.35^4 = 12 mJ

Henrik's test setup (originally placed just outside the MPA):


Image Added

Fig 2: Test setup for new RP-attenuator. Was installed just outside MPA.Image Removed

Reflectivity tests

...