  • Click "Stop" in the "Image Collection" section of the main IOC window
  • Make sure the "Readout Mode" (middle of the IOC sub panel) in the "Collect" section is set to "Image" and NOT "FVB" (full vertical binning)
  • caput RIX:DIR:CAM:01:AndorIsolatedCropMode Enable
  • Set the roi to 90x90 on the main screen in the "Acquisition Settings" section (other settings may work at lower rates. need to ask Dan).
    • NOTE: I don't understand why, but I appear to have to enter the "Y" region size before I enter the "X" region size.
  • Set the "Cam Trig Event Code" to the appropriate value in the "Timestamps" section of the main IOC window
  • Click "Start" in the "image Collection" section of the main IOC window
  • Verify that "Capture Rate" in the "Image Collection" section of the main IOC panel is what you expect

...