run166 event 100 - norman, mean ~ 0 ADU
run 166 event 92 - outlier, mean ~ -25 ADU
Run 167 presumably signal but not so much intensity events 297-300
...
- lcls2/psana/psana/detector/test_01_epix10ka_raw_calib_image.py image -e ueddaq02 -d epixquad -r167 -Sraw-peds -J1 -K295 -N297 -o 2021-03-08-figs/r167-raw-peds-signal-normal-ev0297
- lcls2/psana/psana/detector/test_01_epix10ka_raw_calib_image.py image -e ueddaq02 -d epixquad -r167 -Sraw-peds -J1 -K295 -N298 -o 2021-03-08-figs/r167-raw-peds-signal-outlier-ev0298
- lcls2/psana/psana/detector/test_01_epix10ka_raw_calib_image.py image -e ueddaq02 -d epixquad -r167 -Sraw-peds -J1 -K295 -N299 -o 2021-03-08-figs/r167-raw-peds-signal-outlier-ev0299
- lcls2/psana/psana/detector/test_01_epix10ka_raw_calib_image.py image -e ueddaq02 -d epixquad -r167 -Sraw-peds -J1 -K295 -N300 -o 2021-03-08-figs/r167-raw-peds-signal-normal-ev0300
run 167 event 297 - normal, mean ~ 0 ADU
run 167 event 298 - outlier, mean ~ -25 ADU
run 167 event 299 - outlier "recovering", mean ~ -4 ADU
run 167 event 300 - normal, mean ~ 0 ADU
Issues
- mean value rising during 1000 events of the run 137 step 2/FL
- pedestals for dark are evaluated correctly
- baseline for signal has an offset ~5ADU relative to dark
- there are outliers in signal too which shift mean ~ -25 ADU
...