Page History
...
epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G175,172
BadAir 6, 21, 150
epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G21,150
good plots
BadWater 6, 168, 191
epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G168,191
good plots
BadAir 6, 23, 159
epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G23,159
good plots
Chi2
Commands to generate plots
bsub -o 2020-09-28-mfxc00118-r0172-offset-log06 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o work -p -C -i6
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-H.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-M.dat
chi2 for panel 2 AHL-H, AHL-L, AML-M, AML-L
chi2 for panel 6 AHL-H, AHL-L, AML-M, AML-L
Summary
- Minor detector damage is observed in a few constant types - in charge injection gains for FL, AHL-L, AML-L
- Charge injection gains in stead of uniform constant gains can be used to eliminate this effect.
...