epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G175,172

Image AddedImage AddedImage Added

Image AddedImage AddedImage Added


BadAir 6, 21, 150

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G21,150

good plots

BadWater 6, 168, 191

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G168,191

good plots

BadAir 6, 23, 159

epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o ./work1 -s5 -i6 -G23,159

good plots

Chi2

Commands to generate plots

bsub -o 2020-09-28-mfxc00118-r0172-offset-log06 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r172 -o work -p -C -i6

plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-H.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AHL-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-L.dat
plims panels/0000000001-0176294657-0218103830-1010951301-0019435010-0000000000-0000000000/gain/epix10ka_0069_20200915184723_mfxc00118_r0172_chi2ci_AML-M.dat

chi2 for panel 2 AHL-H, AHL-L, AML-M, AML-L


Image AddedImage AddedImage AddedImage Added

chi2 for panel 6  AHL-H, AHL-L, AML-M, AML-L

Image AddedImage AddedImage AddedImage Added

Summary

  • Minor detector damage is observed in a few constant types - in charge injection gains for FL, AHL-L, AML-L
  • Charge injection gains in stead of uniform constant gains can be used to eliminate this effect.

...