InstructionsAction Fields
Go to Pulsed LLRF displayOpen embedded or start a script?
Set the drive level to 15% and RF On Time to 70 ms. Push "Go." Waveforms will appear and display the calculated cavity voltage based on the Loaded Q calculated above. 
Verify the radiation monitor is logging for the cavity. 
Slowly increase the Drive Level in 5% steps manually (clicking "Go" after each change) until the cavity amplitude reaches AUSE_SRF or the radiation level exceeds 50 MR/hr (0.5 mSv/hr) 
Record the maximum Drive Level usedPush button to store in python variable.

If the cavity faults before reaching 16.6 MV or the usable amplitdue:

  1. Check the radiation monitor for excursions
  2. Check the striptools of temps and vacuums for excursions
  3. Push the "Interlocks" button at the top and record which interlock(s) faulted
  4. Clear faults
  5. Retest the cavity up to 0.5Mv below where it tripped. Continue to lower the maximum value twice more if faulting continues
  6. After it has tripped three times, make note in an elog entry, including why it is faulting, and move on to the next cavity.
Maybe some buttons to help here or at least show the interlocks
Lower the Drive Level to 20% and turn off the RF drive.