- Configure the main registers of the SMURF board (to be detailed)
- Set the bias of the cryogenic amplifiers. The bias are provided by the manufacturer.
- Find the resonator frequencies -
- Each multiplexed line must be scanned in frequency. The resonators are situated where the amplitude of the signal drops and there is a phase discontinuity. The phase discontinuity is used to determine the most precise frequency of the resonators. If 2 resonators are too close, one needs to be turned off to avoid crosstalk.
- SetupNotches - Find the transformation η(I,Q) = (I', Q') where η is a complex tranformationtransformation. We want to move the operation point so that we are at the (I', Q') = (1,0) point and that all variations maximized their projection on Q'.
- Setup the flux ramp
- Flux amplitude
- Ramp amplitude