Configuration Steps

The following are an overview of the main configuration steps for the TES-SMURF detector to go from turn the system on and start collecting data.

To be completed as more information becomes available.

  1. Configure the main registers of the SMURF board (to be detailed)
  2. Set the bias of the cryogenic amplifiers. The bias are provided by the manufacturer. 
  3. Find the resonator frequencies - 
    1. Each multiplexed line must be scanned in frequency. The resonators are situated where the amplitude of the signal drops and there is a phase discontinuity. The phase discontinuity is used to determine the most precise frequency of the resonators. If 2 resonators are too close, one needs to be turned off to avoid crosstalk.

 