Configuration Steps
The following are an overview of the main configuration steps for the TES-SMURF detector to go from turn the system on and start collecting data.
To be completed as more information becomes available.
- Configure the main registers of the SMURF board (to be detailed)
- Set the bias of the cryogenic amplifiers. The bias are provided by the manufacturer.
- Find the resonator frequencies -
- Each multiplexed line must be scanned in frequency. The resonators are situated where the amplitude of the signal drops and there is a phase discontinuity. The phase discontinuity is used to determine the most precise frequency of the resonators. If 2 resonators are too close, one needs to be turned off to avoid crosstalk.