14:20  Calibration for FEI4 telescope progressed with module 2 recovered, but some other modules showing strange results. Noise scans with 4 modules crashes calibserver. This was eventually overcome and a full set of calibrations of all 5 modules stored in their own directories modconfigs/lbnl_cosmic/mod* (*= 0-5). cfg #27  was the eventual state reached that should be used for default running.     

14:25  Ended Run 428 for access. RD53A timing setup adustments, Added a Lemo-T to the TLU beam trigger input to send a signal to HSIO2 Ext trigger LEMO port to allow cosmic telescope to be driven by the beam trigger independently with cosmicGui. Also took out the lemo on first module's HitOR just in case that might be a noise source.

...

16:26: Started Run 441 with both RD53a with two FE-ID set distinctively. Offline checks confirmed hits possition correlation between RD53a and MIMOSA.

16:50  16:50  Tried to run Cosmic Telescope with cosmicGui withe "scintillator trigger" being the beam trigger over LEMO, but no triggers seen. Upon checking Event Log from Sep/2016 session, it was noted there that the beam trigger came over with BNC cable was NIM. HSIO2 LEMO trigger is expecting TTL. 

17:56  End Run 447 to request for access. Changed the RD53a scope monitor scale and brought the LEMO T of beam trigger to an NIM->TTL module and led the TTL LEMO to HSIO-2 TRIG1 port. Also replaced the data/cmd cable for cosmic telescope module 2 which has somewhat ragged noise.        

19:18  Started Run 456 with beam trigger working again. We followed the written procedure of start_1 and found that only got one RD53a (#SPEC1SPEC#1) going. Texted Timon that we need to run start_0 in addition to bring up SPEC#0. 

19:25  Ended 456 to start Run 457 with both RD53a.  

17:55  Tried to run cosmicGui but couldn't even get the cyclic triggers to run. Looking at the date of the cosmicGui image, it was from Sep/10 when Matthias first ported a working rceProducer with EUDAQ-1.7, but the CosmiGui code with EUDAQ dependence was surely not updated yet, as we knew the version running in the lab needed a major overhaul for this a few weeks ago. This has to be checked with Martin/Matthias to revisit tomorrow.   

20:30  Checked the calibration for cosmic telescope module 2 (since cable was changed). This is should a more uniform noise and the jagged shoulder disappeared. Calbration _56 looks like a good default for module 2. 

20:34  Run 458 ended after reaching 10K events and auto started Run 459.