  1. Install a fast diode and mirror (roughly at 45 degrees with respect to the beam) on the SC1 sample_x stage (CXI:SC1:MMS:02.RBV)
    1. The fast diode has a small area and is therefore difficult to align. Ideally, a Y stage (PP-30 piezo) is used to allow remote motion of the diode to align to the laser and x-rays.
    2. Due to the limited travel range of the long X-stage in SC1, the diode and mirror must be installed as far as reasonable towards the -X side, overhanging away from the carriage
  2. Roughly align the fast diode and mirror using the reference laser at the time tool (-4 mm in position but this will soon be changed to 0 mm.   If the "out" position is 92 mm, then -4 mm, if 96 mm, then 0 mm).
  3. Install a regular diode connected to the Acqiris on the sample chambe door and aligned the reflected laser beam on the mirror onto that diode.
    1. Repeat with the fs laser and maximize the acqiris signal by moving the diode on the chamber door.
  4. Insert a nozzle rod with both a clear YAG and a frosted YAG. 
    1. At CXI, the clear YAGs are typically thinner, around 20 um, whereas the frosted YAGs are thicker, around 50 um
  5. Determine the position of the X-ray spot at the IP using the clear YAG and the beam with 1x10-4 Transmission
    1. (zoomed in view)

...