  • CLM 1 (cameralink medium mode)
  • SVM 1 (test pattern ramp).  NOTE: with the test pattern, the camera always seems to read out at the internal line rate, even in external trigger mode.  To have external triggering controlled only by CC1, need to set SVM to 0.
  • SSF 1 (software trigger rate 1Hz, although seems to read back as 6Hz? and 2 reads back as 12Hz?)
  • SEM 0 (Exposure Mode: internally controlled exposure time)
  • SET 20000 (Exposure Time: ns exposure)
  • STM n (Trigger Mode: 0 for internal, 1 for external ... we're guessing there is a typo in the manual where the values should start from 0)
  • SPF 2 (turns to 12 bit. 0 is 8 bit)

Now enable triggers:

  • in "Variables" tab set DataEn for channel A to true
  • Framecount field should increment at 6Hz
  • Dropcount field counts "3 channels misaligned" errors, and so should stay at 0

...