Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The common-mode methods still correlate, but the default method perhaps over-subtracts because of leakage into the zero-photon peak from neighbor-pixels of real photons.

Crystallography Case

Here are the common mode corrections applied to a crystal diffraction pattern with a strong water ring using algorithms 1 vs 5. Algorithm 5 seems to perform better in this case.

Image Added

Common mode histogram using algorithm 1

Image Added

Common mode histogram using algorithm 5

Summary for CSPAD

Common mode correction for CSPAD works with algorithm #1 and shows significant effect.

...