Page History
...
Code Block |
---|
2015-04-02 13:40:57 (info) CalibManager.GUIMetrology: Quality check in X-Y plane: segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 0 segm: 0 20981 20978 -33 -36 -43703 -43708 -17 -22 0.04523 48482 48477 5 3 5 segm: 1 20978 20974 -152 -156 -43707 -43702 -73 -68 0.20189 48474 48480 -6 4 -5 2015-04-02 13:40:57 (info) CalibManager.GUIMetrology: Quality check in Z: segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) Quad 0 segm: 0 20979 -43705 43705 20979 -1 31 -37 -5 15 -21 -32 -32 -21 15 -0.02753 0.04097 32.009 segm: 1 20976 -43704 43704 20976 -28 -38 5 -5 -33 0 10 10 0 -33 0.00000 -0.09014 -10.007 |
Calibration
...
file
/alignment/cspad2x2//reg/g/psdm/detector
calib-cspad2x2-03-2015-04-02/
calib/CsPad2x2::CalibV1/MecTargetChamber.0:Cspad2x2.3/
geometry/0-end.data
center/0-end.data
tilt/0-end.data
Calibration files
In archive
- All files for CSPAD2x2 alignment are stored under directory
/alignment/cspad2x2//reg/g/psdm/detector
- Files named as
*-Metrology.xlsx
and*-Metrology.txt
are original metrology and its text version. - Calibration files are located under
/reg/g/psdm/detector/alignment/cspad2x2/calib/
directory. - Calibration files can be deployed in the experimental calib store using file manager in
Calibration Management Tool
...
Code Block |
---|
/reg/d/psdm/mec/mec73313/calib/CsPad2x2::CalibV1/ MecTargetChamber.0:Cspad2x2.1/<type><calib-type>/0-end.data ... MecTargetChamber.0:Cspad2x2.1/<type><calib-type>/0-end.data |
where <type> stands for geometry, tilt, and center.
Known detectors
on 20142016-0706-0415:
Code Block |
---|
0 : CxiDg2.0:Cspad2x2.0 1 : CxiDg2.0:Cspad2x2.1 2 : CxiSc1CxiDg3.0:Cspad2x2.0 3 : CxiSc2CxiSc1.0:Cspad2x2.0 4 : CxiSc2.0:Cspad2x2.10 5 : CxiSc2.0:Cspad2x2.21 6 : CxiSc2.0:Cspad2x2.32 7 : CxiSc2.0:Cspad2x2.43 8 : CxiSc2.0:Cspad2x2.54 9 : CxiSc2.0:Cspad2x2.65 10 : CxiSc2.0:Cspad2x2.76 11 : MecEndstationCxiSc2.0:Cspad2x2.6 127 12 : CxiSc2.0:Cspad2x2.8 13 : CxiSc2.0:Cspad2x2.9 14 : MecEndstation.0:Cspad2x2.6 15 : MecTargetChamber.0:Cspad2x2.0 13 16 : MecTargetChamber.0:Cspad2x2.1 14 17 : MecTargetChamber.0:Cspad2x2.2 15 18 : MecTargetChamber.0:Cspad2x2.3 16 19 : MecTargetChamber.0:Cspad2x2.4 17 20 : MecTargetChamber.0:Cspad2x2.5 18 21 : SxrBeamline.0:Cspad2x2.2 19 22 : SxrBeamline.0:Cspad2x2.3 20 23 : XcsEndstation.0:Cspad2x2.0 21 24 : XcsEndstation.0:Cspad2x2.1 22 25 : XcsEndstation.0:Cspad2x2.2 26 : XcsEndstation.0:Cspad2x2.3 27 : XcsEndstation.0:Cspad2x2.4 28 : XppGon.0:Cspad2x2.0 23.0 29 : XppGon.0:Cspad2x2.1 30 : XppGon.0:Cspad2x2.12 24 31 : XppGon.0:Cspad2x2.23 25 32 : XppGon.0:Cspad2x2.34 |
Software
Access interface to any detector data and calibration parameters is implemented in Python Package Detector.
...
- Package CSPadPixCoords generates and returns the CSPad2x2 pixel coordinates, taking into account geometry alignment parameters, if available in the expected place under /reg/d/psdm/<instrument>/<experiment>/calib directory.
- Module CSPadPixCoords::CSPad2x2ImageProducer produces CSPad2x2 image, accounting for correct geometry, if available.
Calibration files
/alignment/cspad2x2//reg/g/psdm/detector
calib-cspad2x2-03-2015-04-02/
calib/CsPad2x2::CalibV1/MecTargetChamber.0:Cspad2x2.3/
center/0-end.data
tilt/0-end.data
Overview
Content Tools