name | description | source | units | nominal value |
---|---|---|---|---|
T | air temp in SVT box | dry air sensor | C | between SVT and room temps |
RH | relative humidity in SVT box | dry air sensor | % | near or below 0 |
dew | dewpoint in SVT box | dry air sensor | C | -50 or below |
fl | flow switch state | flow switch | boolean | 0 (flow above setpoint) |
fS | FEB chiller serial link | serial port | boolean | 1 (OK) |
sS | SVT chiller serial link | serial port | boolean | 1 (OK) |
fT | FEB chiller temp | FEB chiller serial link | C | setpoint +/- 0.1 |
fSP | FEB chiller setpoint | FEB chiller serial link | C | 20 - 30 |
fFl | FEB chiller flow | FEB chiller serial link | gpm | 0.30 - 1.0 (this flow meter is not very reliable) |
fF | FEB chiller fault code | FEB chiller serial link | 0 (no errors) | |
sC | SVT chiller state | SVT chiller serial link | boolean | 1 (running) |
sT | SVT chiller temp | SVT chiller serial link | C | setpoint +/- 0.2 |
sSP | SVT chiller setpoint | SVT chiller serial link | C | -20 - 20 |
sPw | SVT chiller power | SVT chiller serial link | % full power | 15 - 30 |
sP | SVT chiller pressure | SVT chiller serial link | bar | 1.5 - 2.0 |
io1 | BNC input voltage 1 | BNC jack | V | 0 - 5.0 |
...