  1. If on, turn off NIM crate to disable interlock (otherwise chiller won't start)
  2. Turn on SVT power supply crate (turn key to "local on"), make sure red "interlock" light comes on (i.e. tripped)
  3. Set flow manifold valves
  4. Start chiller (power button on front panel)
  5. Turn on NIM crate, make sure Arduino's green "ON" light and flow switch's red "flow above threshold" light turn on, SVT power supply's red interlock light turns off
  6. Do stuff to start power and DAQ

SVT Lab Layout

Top half:

  • physically on away side from chiller
  • cooling loop on away side from chiller
  • flange with long cable outside thermal box

Systems

Power

CAEN SY1527

Serial console

Serial port is connected to Windows desktop; run HyperTerminal (Start/Prorams/Accessories/Communication/HyperTerminal) and open the "CAEN SY1527" shortcut (set to COM1, 9600 baud, hardware flow control).

Instructions

In the current setup:

Slot

PS

Type

0

A509

DVDD for all hybrids

3

A509H

AVDD, V125 and bias

6

A509H

AVDD, V125 and bias

9

A509H

AVDD, V125 and bias

Wiki Markup
*\[Update with correct slot/ch assignment valid for group C\]*
Mapping between the SY527 slots and channels to hybrids can be found: half-module assignments - Power supply assignments updated.pdf .

Cooling

Chiller

Polyscience 6506T

...